Ambient Conditions Change Baseline 0 on an IR Breath instrument - Disappearing Alcohol

Writer's picture: Stephen BissStephen Biss

Updated: Sep 13, 2022

Experiment studying the phenomenon of disappearing alcohol


It is easy to replicate this experiment on an Intoxilyzer® 5000. It is more difficult to replicate this experiment on an Intoxilyzer® 8000 because the 8000 has a very weak pump. The phenomenon applies, however, to both instruments because the baseline zero is always floating.


Apparatus:


Intoxilyzer® 5000

Wet-bath simulator with 100 target value solution at 34.0 C

Wet-bath simulator with 50 target value solution at 34.0 C


This video displays the function of and relationship between the air blank and ambient fail systems on an IR evidentiary breath instrument such as the Intoxilyzer® 5000C (or 5000 as in this video) and the Intoxilyzer® 8000C (or 8000). In this experiment the room's ambient conditions have been artificially elevated to 50 mg / 100 mls by attaching a second wet-bath simulator (at 50 mg/ 100 mls and 34C) to the breath tube inlet.


Although the 5000 instrument is programmed to flag "Ambient Fail" at 20 mg/100 mls, no such error message is produced. This is because the instrument can only compare an unacceptable ambient situation ( 50 mg/100 mls) against the baseline established when the instrument was first powered on (50 mg / 100 mls.) Note that the instrument is powered up (cold boot) with the 50 mg/100mls simulator output attached to the breath tube.


The instrument erroneously considers the 50 mg /100 mls as 00 mg /100 mls. Any subsequent breath or cal. check readings will be affected by this baseline. The baseline zero, however, shifts up and down relative to each air blank.


Why is this significant to defence counsel?


The real danger of this phenomenon is that calibration checks become not reliable - and in this example the 100 mg /100 mls alcohol standard produces a calibration check of 51 mg / 100 mls. COBRA data from the weeks before and after a particular subject test may be very useful in flagging unusually low cal checks that possibly are being caused by ambient ethanol. Alternatively, another chemical interferent in the room air (e.g. hand cleaners, paint, cleaning fluids) may do the same thing as ambient ethanol. Unreliable cal checks do not cause erroneous subject test results but they do compromise the scientific reliability of the subject tests, and should compromise the Crown's ability to rely on the 320.31(1)(a) presumption of accuracy.


Recent Posts

See All

Comentários


If you are a member of the public, please don't attempt to use what you see or read at this site in Court. It is not evidence. The author is not a scientist. The author has a great deal of experience in cross-examining scientists about these issues, but the author is not a scientist. Hire a criminal lawyer in private practice in Ontario. Your lawyer can retain an expert. The author is a retired lawyer, not a lawyer in private practice. Read the statement of the purpose of this web site below.

© 2025 Allbiss Lawdata Ltd.

This site has been built by Allbiss Lawdata Ltd. All rights reserved. This is not a government web site.

For more information respecting this database or to report misuse contact: Allbiss Lawdata Ltd., Mississauga, Ontario, Canada, 905-273-3322. The author and the participants make no representation or warranty  whatsoever as to the authenticity and reliability of the information contained herein.  WARNING: All information contained herein is provided  for the purpose of discussion and peer review only and should not be construed as formal legal advice. The authors disclaim any and all liability resulting from reliance upon such information. You are strongly encouraged to seek professional legal advice before relying upon any of the information contained herein. Legal advice should be sought directly from a properly retained lawyer or attorney. 

WARNING: Please do not attempt to use any text, image, or video that you see on this site in Court. These comments, images, and videos are NOT EVIDENCE. The Courts will need to hear evidence from a properly qualified expert. The author is not a scientist. The author is not an expert. These pages exist to promote discussion among defence lawyers.

Intoxilyzer®  is a registered trademark of CMI, Inc. The Intoxilyzer® 5000C is an "approved instrument" in Canada.

Breathalyzer® is a registered trademark of Draeger Safety, Inc., Breathalyzer Division. The owner of the trademark is Robert F. Borkenstein and Draeger Safety, Inc. has leased the exclusive rights of use from him. The Breathalyzer® 900 and Breathalyzer® 900A were "approved instruments" in Canada.

Alcotest® is a registered trademark of Draeger Safety, Inc. The Alcotest® 7410 GLC and 6810 are each an "approved screening device" in Canada.

Datamaster®  is a registered trademark of National Patent Analytical Systems, Inc.  The BAC Datamaster® C  is an "approved instrument" in Canada.

bottom of page