Know the Simulator Probes

Updated: Nov 13, 2022

Tip 24: Knowledge of #duisimulator probes assists understanding of function & malfunction of thermistor, heater, and thermometer(s).


Many of our trials involve events that happened in Canadian winter conditions. Running evidentiary breath testing of subjects from a breath truck can be challenging for a qualified technician. Defence lawyers need to pay attention to whether or not the wet-bath simulator used in winter in a breath truck was being used properly. Pay close attention to:


  1. how each of the probes in the simulator respond to temperature variations (this blog entry)

  2. simulator heater response time (see another blog entry)

  3. manufacturer's specifications minimum operating temperatures - use of simulator in winter conditions (see another blog entry)

The Guth 2100 Dispersion Tube bubbles air pumped from the approved instrument during the cal check sequence. That air passes through the liquid alcohol standard and pushes alcohol vapour air up through the top of the simulator and is sucked into the approved instrument sample chamber. Hopefully, if there is a tight seal, correct temperature, and equilibrium, the contents of that alcohol standard air reflect the target value of the alcohol standard, usually 100 mg / 100 mls.

The Guth 2100 Agitator (Propellor) is continuously turning to ensure that the contents of the liquid below and the air above are all at the same temperature and at are their respective target concentrations (Henry's Law). A simulator with a propellor that is unreliable may not be maintaining the equilibrium.

The Heating Element turns on and off to heat the liquid contents of the jar to maintain 34.00 ± .05° C.. The heating element is controlled by the heater control that gets its information about current temperature from the micro controller attached to the Temperature Sensor probe which contains a thermistor (similar to a thermostat).


Guth 2100 probes
Probes on the author's Guth 2100

It is the responsibility of the Qualified Technician to record the temperature of the alcohol standard in the #duisimulator system at the time of any cal check to ensure that the alcohol standard vapour being pumped into the approved instrument sample chamber is at 34.0 ± .2° C (Henry's Law reflected in ATC Recommendations). If the temperature of the vapour pumped into the sample chamber is not within that temperature range, then the alcohol standard is NOT RELIABLE, and the subject breath test is NOT RELIABLE.

A good way to check and record that temperature would be for the QT to observe the Optional NIST Mercury Thermometer during the cal check and key that information into the approved instrument or record it in his or her notes.

The usual way, however, is for the QT to observe the temperature on the digital display or the RS232 interface display and report/record that temperature. Quaere: Where is that temperature information coming from? What temperature sensor sends the temperature information to the digital display or the RS232 interface? Is the same "precision thermistor" that controls the heater, also the probe that is the source of temperature information to the QT via the digital display or RS232 interface?

Please study the diagram at page 30 of 238 of the 2013 CFS Intoxilyzer 8000C Training Aid. With respect, the 2013 diagram of the "Guth Laboratories Model 2100 & Guth Laboratories Model 10-4D" is incorrect in its placement of the probes. The diagram at page 5-11 of the 2011 Training Aid is better. Please also see page 3 of the Guth Laboratories Inc. : "Model 2100 Simulator Operation Manual".

Comentários


If you are a member of the public, please don't attempt to use what you see or read at this site in Court. It is not evidence. The author is not a scientist. The author has a great deal of experience in cross-examining scientists about these issues, but the author is not a scientist. Hire a criminal lawyer in private practice in Ontario. Your lawyer can retain an expert. The author is a retired lawyer, not a lawyer in private practice. Read the statement of the purpose of this web site below.

© 2025 Allbiss Lawdata Ltd.

This site has been built by Allbiss Lawdata Ltd. All rights reserved. This is not a government web site.

For more information respecting this database or to report misuse contact: Allbiss Lawdata Ltd., Mississauga, Ontario, Canada, 905-273-3322. The author and the participants make no representation or warranty  whatsoever as to the authenticity and reliability of the information contained herein.  WARNING: All information contained herein is provided  for the purpose of discussion and peer review only and should not be construed as formal legal advice. The authors disclaim any and all liability resulting from reliance upon such information. You are strongly encouraged to seek professional legal advice before relying upon any of the information contained herein. Legal advice should be sought directly from a properly retained lawyer or attorney. 

WARNING: Please do not attempt to use any text, image, or video that you see on this site in Court. These comments, images, and videos are NOT EVIDENCE. The Courts will need to hear evidence from a properly qualified expert. The author is not a scientist. The author is not an expert. These pages exist to promote discussion among defence lawyers.

Intoxilyzer®  is a registered trademark of CMI, Inc. The Intoxilyzer® 5000C is an "approved instrument" in Canada.

Breathalyzer® is a registered trademark of Draeger Safety, Inc., Breathalyzer Division. The owner of the trademark is Robert F. Borkenstein and Draeger Safety, Inc. has leased the exclusive rights of use from him. The Breathalyzer® 900 and Breathalyzer® 900A were "approved instruments" in Canada.

Alcotest® is a registered trademark of Draeger Safety, Inc. The Alcotest® 7410 GLC and 6810 are each an "approved screening device" in Canada.

Datamaster®  is a registered trademark of National Patent Analytical Systems, Inc.  The BAC Datamaster® C  is an "approved instrument" in Canada.

bottom of page