top of page

lawyers.ca duimetrology.com
905-273-3322 or 1-877-273-3322
All Videos in the DUI Video Collection
Links to DUI Videos on Specific Topics:
Filter list for Radio Frequency Interference Videos only
Filter list for Chemical Interferent Videos Only
Filter List for Mouth Alcohol Bias Videos Only
Video 52
Cell Phone RFI Error Interferent 64 Series
This is an example of a false Error Message. The instrument indicates "Interferent" i.e. a chemical interferent when the cause is radio frequency interference (RFI). Error messages are not always reliable to identify the correct error.
ACACA calibration checks using 100 solution, first cal. check 102, second cal. check indicates "Interferent" then 106 when exposed to cell phone RFI
Ontario
Intoxilyzer 5000
RFI | Interferent | MouthAlcohol | Maintenance | Keyboard | Errors | Reliability | SourceCode |
---|---|---|---|---|---|---|---|
Y | Y |
bottom of page