lawyers.ca duimetrology.com
905-273-3322 or 1-877-273-3322
All Videos in the DUI Video Collection
Links to DUI Videos on Specific Topics:
Filter list for Radio Frequency Interference Videos only
Filter list for Chemical Interferent Videos Only
Filter List for Mouth Alcohol Bias Videos Only
Video 17
RAMCheck2 0019 RFI Causing 011 to 014, 010, 007, 000, 022 to 027, 022, 024 to 029, can't clear AIR BLANK, 000
This is an example of RFI causing erroneous BAC analysis indications on an evidentiary test instrument without triggering an "RFI Interference" error message. Please notice the "purge fail" phenomenon when the instrument cannot clear the 029 result (zero has been reset too low by the RFI).
Intoxilyzer 5000 in need of maintenance. Intoxilyzer® is a registered trademark of CMI, Inc. Instrument in ACACACACACA mode. Print inhibit on. No simulator, no source of ethanol or interferent. RFI from two Cobra walkie-talkies Channel 3 00 used, one transmitting, another receiving, so you can hear when "Talk" is pressed. The Intoxilyzer® 5000C is an "approved instrument" in Canada.
Ontario
Intoxilyzer 5000 64 series upgraded to 66
RFI | Interferent | MouthAlcohol | Maintenance | Keyboard | Errors | Reliability | SourceCode |
---|---|---|---|---|---|---|---|
Y |